Brewster Angle Microscope (BAM) EP3
The Accurion EP3-BAM is used to image thin films on solid or liquid substrates. To obtain image contrast, the BAM uses the principle that at a certain incidence (Brewster) angle the p-polarized laser beam is not reflected at all by the substrate-air interface. The Brewster angle depends only on the refractive index (RI) of the substrate and is equal to 53° in the case of a water substrate. At this incidence angle, nano-thick molecular and/or nanoparticle films that are spread at the water surface will modify locally the RI value, thus yielding a non-zero reflection coefficient and providing image contrast.
Instrument parameters:
-
- Light source: laser unit λ 532nm
- Imaging system CCD camera 768 x 572px
- Imaging optics:
- 10X CF Plan Nikon Objective: NA/0.21; Working distance/20.3mm
- 20X CF Plan Nikon Objective: NA/0.21; Working distance/20.3mm
-
- Motorized goniometer:
- Angle of incidence (AOI) range: 40-90°
- Angle resolution: 0.001°
- Absolute angle accuracy: 0.01°
- Motorized goniometer: